Did you know...

Each X Ray machine has to be tested to measure the amount of radiation it emits once installed.

As part of the Mercia Dental service, our engineers perform thorough NRPB tests on each machine after installation. These tests are then recorded and the results are sent to the NRPB.

Finance

 

Technical details



Satelec X-Mind DC



Satelec X-Mind DCSatelec-x-mind-controller

Technical details:

Classification: Electromedical equipment, Class 1 type B
Supply voltage: 230V~ - monophase 50/60Hz
Power absorption at 230V: 1.4kVA
Resistance of the line: 0.5Ω
X-ray tube: New Toshiba DG 073B DC tube
High voltage: 60-70kV
Anode current: 4-8mA
Focal spot: 0.7mm
Total filtration: Equivalent to 2mm Al at 70kV
Tube inherent filtration: Equivalent to 0.8mm Al at 70kV
Leakage radiation: Less than 0.25mGy/hour maximum at a distance
of 1m from focal spot
Long cone: Focus to cylinder tip distance = 31cm (12”)
Timer: Exposure time can be set from 0.02 to 3.2 seconds
X-ray emission control: “Dead man” switch with 3m spiral cable
Total weight: 25kg (weight of the head: 5.5kg)


Manufactured in compliance with currently applicable regulations and standards (EC Directive 93/42/EEC)